News | Angiographic Catheter | October 14, 2015

Cook Medical Expands Global Recall of Select Beacon Tip Angiographic Catheter Sizes

Recall is an expansion of voluntary lot-specific recall issued in July

Cook Medical, expanded recall, October 2015, Beacon Tip Angiographic Catheters

Image courtesy of Cook Medical


October 14, 2015 — On Oct. 7, 2015, Cook Medical initiated a voluntary recall for select sizes of Beacon Tip Angiographic Catheters. This recall includes all lots of these select sizes of the Beacon Tip Angiographic Catheters. This recall is an expansion of the voluntary lot-specific recall issued on July 2, 2015. The products include specific versions of the Torcon NB Advantage Beacon Tip Catheters (catalog prefix HNBR4.0, HNBR4.1 and only HNBR5.0 with the RUC suffix), Royal Flush Plus Beacon Tip High-Flow Catheters (catalog prefix HNR4.0), and Slip-Cath Beacon Tip Hydrophilic Catheters (catalog prefix SCBR4.0, SCBR4.1, and only SCBR5.0 with the RUC suffix), and Shuttle Select Slip-Cath catheters (catalog prefix SCBR4.5).

The list of lot numbers affected by the recall can be found here.

The catheters have been found to exhibit tip splitting or separation, which has resulted in 42 Medical Device Reports. Tip splitting has the potential to lead to loss of device function. Tip separation may require medical intervention to retrieve a separated segment or, if not retrieved, has the potential to occlude blood flow to end organs.

The Beacon Tip Angiographic Catheters in this recall were distributed globally between September 2012 and September 2015. Product can be identified by the part number provided on the outer package product label. The part numbers for products subject to this recall are those with catalog number prefixes and suffixes listed in the initial paragraph above. All lots currently in distribution are affected.

Cook Medical has notified its customers and distributors by recall notification letters. The letters requested that all customers and distributors quarantine and discontinue use of all potentially affected units and return the affected product to the company as soon as possible for credit.

The U.S. Food and Drug Administration (FDA) and other regulatory agencies around the world have been notified of this action.

Consumers with questions may contact Cook Medical Customer Relations at 1-800-457-4500 or 1-812-339- 2235, Monday through Friday, between 7:30 a.m. and 5:00 p.m. Eastern time or by email at [email protected].

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

For more information: www.cookmedical.com


Related Content

News | Cath Lab

December 20, 2023 — Jason R. McCarthy, Ph.D., associate professor of biomedical research and translational medicine and ...

Home December 20, 2023
Home
News | Cath Lab

October 26, 2023 — Royal Philips, a global leader in health technology, announced the latest results demonstrating the ...

Home October 26, 2023
Home
News | Cath Lab

October 25, 2023 — Shockwave Medical, Inc., a pioneer in the development and commercialization of transformational ...

Home October 25, 2023
Home
News | Cath Lab

October 20, 2023 — Over the coming days, Philips will be presenting its latest solutions in cardiology and new late ...

Home October 20, 2023
Home
News | Cath Lab

October 16, 2023 — GE HealthCare (Nasdaq: GEHC) announced US FDA 510(k) clearance of Allia IGS Pulse - the latest ...

Home October 16, 2023
Home
News | Cath Lab

October 16, 2023 — Shimadzu Medical Systems USA, a subsidiary of Shimadzu Corporation, announced the first U.S ...

Home October 16, 2023
Home
News | Cath Lab

September 13, 2023 — A diagnostic test, first offered in the United States at University Hospitals (UH) Harrington Heart ...

Home September 13, 2023
Home
News | Cath Lab

August 2, 2023 — Teleflex Incorporated, a leading global provider of medical technologies, announced the U.S Food and ...

Home August 02, 2023
Home
News | Cath Lab

July 13, 2023 — Mount Sinai Queens announced the opening of a new cardiac catheterization lab that will provide rapid ...

Home July 13, 2023
Home
News | Cath Lab

June 21, 2023 — Royal Philips, a global leader in health technology, announced it has teamed up with BIOTRONIK (Lake ...

Home June 21, 2023
Home
Subscribe Now