Feature | July 29, 2013

GE Healthcare Recalls Many of its Nuclear Imaging Systems

Patient's death from falling scanner component spurs total safety recall


July 29, 2013 — GE Healthcare has initiated a Class I recall of most of its SPECT nuclear imaging systems and called for its systems to not be operated until the vendor can inspect the scanner. The recall follows the death of a patient after a portion of the scanner fell onto them during the scan on an Infinia Hawkeye 4 system at a VA medical center facility in New York.

On June 17, GE sent an urgent medical device correction letter to all affected customers. The letter identified the affected product, and recommended that qualified service personnel maintain the equipment and that preventative maintenance procedures are executed according to labeling. In addition, users were told that safety chapter sections should be re-reviewed with personnel to ensure proper operation of the equipment.

On July 3, GE notified hospitals again via an urgent medical device recall letter and follow-up telephone calls, notifying that it was recalling several more of its nuclear imaging systems. The company said serious injuries or death could occur due to the failure mode associated with the recall. In the second notification, GE included most of its nuclear medicine systems, including the Infinia nuclear medicine systems, VG and VG Hawkeye nuclear medicine systems, Helix nuclear medicine systems, Brivo NM615, Discovery NM630, Optima NM/CT640 and the Discovery NM/CT670.

Healthcare facilities are instructed to cease use of the above mentioned nuclear imaging systems until a GE Healthcare field engineer is able to do a complete inspection of the system and perform any necessary repairs at no cost. GE Healthcare representatives will contact hospitals to arrange for the inspection. If an issue with the support mechanism fasteners is found on a system, the GE field engineer will coordinate the replacement of impacted parts in the gantry to ensure the system is operating safely and meets all specifications.

The weight of nuclear imaging cameras is around 1,300 pounds due to the cameras being lead lined and additional lead in the construction of the internal collimation system.  This significant weight poses design challenges in these systems because of the fatigue on the supports. This is an issue faced by all vendors' system and is not unique to GE. 
 

Investigation of the Accident

In its July 3 recall letter, GE said it was informed of the patient’s death on June 5. GE Healthcare responded and is working with the facility and all appropriate government agencies to complete a thorough investigation.

The company was permitted access to the equipment on June 17. GE Healthcare determined the bolts securing the camera to the gantry were loose, thereby stressing the support mechanism and resulting in the incident. The safety concern is related to a potential patient entrapment or crush hazard if the camera falls during a patient exam, the company said.

Because of the similarities in the design of support mechanisms across many GE products, the company included in its June 17 recall all types and manufacturing dates of Infinia and Infinia Hawkeye nuclear medicine systems; all types and manufacturing dates of VG and VG Hawkeye nuclear medicine systems; and all types and manufacturing dates of Helix nuclear medicine systems by Elscint Ltd. In the July 3 letter the company expanded this recall to include the Brivo NM615, Discovery NM630, Optima NM/CT640 and the Discovery NM/CT670 due to similarities in basic mechanical concepts between these devices and the Infinia systems.

As of Aug. 1, GE reported it had nearly complete its inspections of its SPECT systems installed globally and that systems requiring repair will be returned to service shortly. 

GE Healthcare representatives are contacting all known medical facilities that have a GE system to arrange for an inspection. For questions about this recall, contact your GE Healthcare service representative at (800) 437-1171.??

Read the July 3 recall letter here.


Related Content

News | Nuclear Imaging

August 3, 2023 — The cardiovascular polypill, developed by the Spanish National Centre for Cardiovascular Research (CNIC ...

Home August 03, 2023
Home
News | Nuclear Imaging

January 5, 2023 — A new study has determined that approximately three percent of all bone scan patients have markers of ...

Home January 05, 2023
Home
News | Nuclear Imaging

April 13, 2022 — The Society of Nuclear Medicine and Molecular Imaging (SNMMI) and the American Society of Nuclear ...

Home April 13, 2022
Home
Videos | Nuclear Imaging

American Society of Nuclear Cardiology (ASNC) President Dennis Calnon, M.D., MASNC, FASE, FSCCT, director of cardiac ...

Home February 01, 2022
Home
Feature | Nuclear Imaging | By Staff of the American Society of Nuclear Cardiology (ASNC)

A year after COVID-19 turned the world upside down, the American Society of Nuclear Cardiology (ASNC) asked members how ...

Home June 02, 2021
Home
News | Nuclear Imaging

April 1, 2021 – The ability to measure myocardial blood flow (MBF) as part of myocardial perfusion imaging (MPI) is one ...

Home April 01, 2021
Home
News | Nuclear Imaging

December 4, 2020 — Spectrum Dynamics Medical, Inc., has received Canadian Medical Device License from Health Canada for ...

Home December 04, 2020
Home
News | Nuclear Imaging

September 30, 2020 — Siemens Healthineers has introduced a new version of its c.cam dedicated cardiac nuclear medicine ...

Home October 07, 2020
Home
Feature | Nuclear Imaging | Dave Fornell, Editor

There were a few key takeaways from the American Society of Nuclear Cardiology (ASNC) 2019 annual meeting in September ...

Home November 22, 2019
Home
Videos | Nuclear Imaging

Rob Beanlands, M.D., FASNC, 2019 American Society of Nuclear Cardiology (ASNC) president, shares a couple trends he sees ...

Home November 07, 2019
Home
Subscribe Now