News | FDA | May 15, 2024

FDA Announces Abbott Recall of HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kit

Abbott recalls HeartMate 3 left ventricular assist system (LVAS) implant kit for risk of blood leakage or air entering system between inflow cannula and apical cuff

Abbott recalls HeartMate 3 left ventricular assist system (LVAS) implant kit for risk of blood leakage or air entering system between inflow cannula and apical cuff

May 15, 2024 — The U.S. Food and Drug Administration (FDA) announced that Abbott is recalling the HeartMate 3 LVAS by updating the instructions for use after complaint reviews identified blood leakage or air entering the seal interface between the left ventricular assist device (LVAD) inflow cannula and the titanium apical cuff. In all reported events, the issue was observed during implantation of the device.

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these affected devices may cause serious injuries, serious health consequences, or death. Please be aware, this recall is a correction, not a product removal.

Recalled Product 

  • Product Name: HeartMate 3 Left Ventricular Assist System Implant Kit
  • Product Codes: See Recall Database Entries:
  • Devices Recalled in the U.S.: 882 
  • Dates Distributed: March 1, 2021 to present
  • Date Initiated by Firm: March 13, 2024

Device Use

The HeartMate 3 Left Ventricular Assist System (LVAS) is used to help the heart pump blood when it is not able to do so effectively on its own. The HeartMate 3 is used for both short- and long-term support in pediatric and adult patients with severe left ventricular heart failure. It can be used while waiting for a heart transplant, to help the heart recover, or as a permanent solution when a transplant isn't an option. 

HeartMate 3 LVAS takes over the pumping function of the left ventricle, the main pumping chamber of the heart. It diverts blood from the weakened left ventricle and propels it into the body's main artery, the aorta, helping to circulate blood throughout the body. The device is used both inside and outside of the hospital.

Reason for Recall  

Abbott is recalling the HeartMate 3 LVAS after complaint reviews identified blood leakage or air entering the seal interface between the left ventricular assist device (LVAD) inflow cannula and the apical cuff. In all reported events, the issue was observed during implantation of the device. 

Blood leakage or air entering the LVAD from this location will impact the integrity of the blood flow and may lead to longer than expected surgery, bleeding (hemorrhage), right heart failure, or air embolism. Use of these devices may cause serious injury or death. 

Abbott reports 81 incidents, including 70 injuries and two deaths related to this issue. 

Who May Be Affected 

  • Patients who need cardiac support using the HeartMate 3 LVAS System. 
  • Health care providers who plan to use the HeartMate 3 LVAS System for their patients. 

What to Do 

On March 20, 2024, Abbott issued an Urgent Medical Device Correction Notification letter to customers recommending the following actions: 

  • Do not return unused product as the product is not being removed from the field. 
  • If blood leak or air entrainment is suspected or observed, follow standard surgical processes and the existing instructions for use (IFU)
    • Residual air must be completely evacuated from the device blood chamber prior to initiating the LVAD support. 
    • Ensure that bleeding is assessed and ensure proper management of hemostasis before closing all wounds. 
    • Use conventional strategies for resolving air leaks or surgical bleeding, including: adjusting the pump position, waiting for the natural tendency of blood to coagulate or upon reversal of anticoagulation, adding surgical materials, and exchanging the apical cuff, the pump, or both. 
    • Always have a complete backup system (implant kit and external components) available on-site and in close proximity during the implantation procedure for use in the event of an emergency. 
  • Distribute the notice to those who need to be aware within the institution. 
  • Complete the acknowledgment form included with the letter and return it to Abbott at MCSHMNotices@abbott.com.

Contact Information 

Customers with questions should contact their local Abbott representative or call 612-346-3514.

Additional Resources 

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program using an online form, regular mail, or FAX.


Related Content

News | FDA

July 7, 2025 — Boston Scientific Corp. has received U.S. Food and Drug Administration (FDA) approval to expand the ...

Home July 08, 2025
Home
News | FDA

June 12, 2025 — Viz.ai recently announced it has received U.S. Food and Drug Administration (FDA) 510(k) clearance for ...

Home June 12, 2025
Home
News | FDA

Nov, 4, 2024 – R3 Vascular Inc. has announced that the U.S. Food and Drug Administration (FDA) granted investigational ...

Home November 07, 2024
Home
News | FDA

July 29, 2024 — CorVascular, a leading producer of peripheral arterial disease (PAD) / peripheral vascular disease (PVD) ...

Home July 29, 2024
Home
News | FDA

July 2, 2024 — The U.S. Food and Drug Administration (FDA) announced that Abbott has issued a correction for its ...

Home July 02, 2024
Home
News | FDA

June 18, 2024 — Elixir Medical has announced the company’s novel bioadaptive implant, DynamX Sirolimus-Eluting Coronary ...

Home June 18, 2024
Home
News | FDA

June 17, 2024 — Elutia Inc., a pioneer in drug-eluting biomatrix products, today announced that its Antibiotic-Eluting ...

Home June 17, 2024
Home
News | FDA

June 13, 2024 — Xeltis, a leading developer of transformative implants that enable the natural creation of living and ...

Home June 13, 2024
Home
News | FDA

June 3, 2024 — Heuron, a specialized medical AI startup focused on brain and neurological disorders based in South Korea ...

Home June 03, 2024
Home
News | FDA

May 17, 2024 — Implicity, a leader in remote patient monitoring and cardiac data management solutions, announced it has ...

Home May 17, 2024
Home
Subscribe Now