Cordis Recalls CROSSOVER Sheath Introducer Due to Fractures

November 4, 2009 – Cordis and FDA notified healthcare professionals this week of a nationwide recall of all lots of the CROSSOVER Sheath Introducer.

The long-coil reinforced, kink-resistant catheter sheath is intended for use in arterial and venous procedures requiring the percutaneous introduction of therapeutic or diagnostic intravascular devices or fluids. The recall was due to stretching or fracture of the sheath during use.

The FDA said in the event of a device fracture, separated segments of the device can embolize downstream in the bloodstream and impede blood flow distal to the point where it lodges, resulting in ischemia or infarct to the distal extremity. Since this device is coil reinforced, any separation of the cannula has the potential to expose portions of the coil creating the potential for vessel dissection or perforation. Unplanned open surgery may be required to remove the retained segments or control bleeding.

For more information: www.fda.gov/Safety/MedWatch


Related Content

Feature | Vascular Access | Dave Fornell, Editor

Vascular access site bleeding is associated with higher complications and mortality rates. For decades femoral access ...

Home June 10, 2020
Home
News | Vascular Access

June 8, 2020 – BD (Becton, Dickinson and Company) launched the Halo One Thin-Walled Guiding Sheath, designed to perform ...

Home June 08, 2020
Home
News | Vascular Access

November 6, 2019 — Saranas Inc. announced completion of the first U.S. commercial case using its Early Bird Bleed ...

Home November 07, 2019
Home
Technology | Vascular Access

September 3, 2019 — New Zealand-based Adept Medical announced the launch of the Antegrade IR Platform. Clinically driven ...

Home September 03, 2019
Home
News | Vascular Access

May 31, 2019 — Terumo Medical Corp. is recalling the SoloPath Balloon Expandable TransFemoral System and Re-Collapsible ...

Home May 31, 2019
Home
News | Vascular Access

October 4, 2017 — TVA Medical Inc. announced that its everlinQ 4 endoAVF System has received CE Mark in the European ...

Home October 04, 2017
Home
News | Vascular Access

March 31, 2017 — The U.S. Food and Drug Administration (FDA) announced it has initiated a Class I recall of the Merit ...

Home March 31, 2017
Home
News | Vascular Access

January 31, 2017 — Bard Peripheral Vascular Inc. is recalling the Halo One Thin-Walled Guiding Sheath because the sheath ...

Home January 31, 2017
Home
Technology | Vascular Access

January 10, 2017 — Teleflex Inc. announced that its Arrow VPS Rhythm Device with optional TipTracker technology has been ...

Home January 10, 2017
Home
News | Vascular Access

September 20, 2016 — Teleflex Inc. recently displayed its vascular access technologies and a new educational platform ...

Home September 20, 2016
Home
Subscribe Now